Individual calibration certificates with expanded measurement uncertainty statements (k=2) per ISO/IEC 17025 framework — for RTD sensors and thermocouple assemblies at customer-specified temperature setpoints. Supplied at time of manufacture or as a standalone calibration service.
NIST-traceable resistance vs. temperature calibration for PT100, PT500, and PT1000 RTD assemblies. Data reported as resistance (Ω) at each setpoint, with deviation from IEC 60751 nominal. Useful for pharmaceutical validation, process qualification, and high-accuracy measurement systems.
EMF vs. temperature calibration for base metal (K, J, T, N, E) and noble metal (R, S, B) thermocouple assemblies per IEC 60584-1. Data reported as EMF (mV) at each setpoint, with deviation from IEC 60584 reference function. Required for AMS 2750F SAT sensor qualification and process validation.
Sets of matched sensors calibrated together for Temperature Uniformity Survey (TUS), process qualification, and validation protocols. Each sensor receives an individual certificate with a common set serial number for batch traceability. Lyophilizer mapping, autoclave validation, and furnace TUS applications.
AMS 2750F requires calibrated TCs for SAT (System Accuracy Test) sensors. NADCAP suppliers require individual cal certs with current calibration dates and traceability chain.
21 CFR Part 11 and EU GMP require documented calibration traceability for all process monitoring instruments used in product manufacturing and storage.
Temperature uniformity mapping and validation protocols require NIST-traceable cal certs with uncertainty statements for each sensor used in IQ/OQ/PQ testing.
ISO 9001:2015 requires measurement equipment to be calibrated at specified intervals against measurement standards traceable to national or international standards.
Research publications require measurement traceability to SI. Test reports must include calibration data and uncertainty. ISO/IEC 17025 accredited lab requirements.
Turbine and boiler temperature monitoring for NERC compliance and insurance requirements. Calibrated sensors with documented traceability for critical protection systems.
Safety Instrumented Systems (SIS) per IEC 61511 require documented instrument calibration and proof testing intervals for sensors in safety instrumented functions (SIF).
HACCP critical control point temperatures require calibrated sensors with documented traceability. FDA inspection may require calibration records for CCP monitoring instruments.
Calibrated against NIST-traceable reference — PT25 capsule SPRT or noble metal TC with current calibration certificate. Traceability chain maintained and documented.
Temperature bath or calibration furnace stabilized at setpoint. Stability criterion: ≤0.05°C drift over 5 minutes before measurement begins.
Reference and DUT measured by calibrated readout instruments (Fluke 1594A or equivalent). Minimum 5 readings per setpoint recorded and averaged.
Combined standard uncertainty calculated per GUM (JCGM 100:2008). Includes reference uncertainty, repeatability, readout resolution, and thermal gradient contributions.
Individual PDF certificate with unique serial number, setpoint data table, correction values, expanded uncertainty (k=2), and QA signature. Issued per AS9100D QMS.
Customer PO number, Thermometrics order number, sensor model, unique serial number, and date of calibration.
Temperature setpoints in °C, measured resistance (RTD in Ω) or EMF (TC in mV), deviation from IEC nominal, and correction factor at each point.
Expanded measurement uncertainty (U) at 95% confidence level (coverage factor k=2) per GUM. Stated at each setpoint or as a single worst-case value.
Statement confirming traceability to NIST through an unbroken chain of comparisons with stated uncertainties. Reference standard ID and calibration date included.
Signed by Thermometrics Quality Assurance. AS9100D document control number. Certificate revision control.
Detailed Type A and Type B uncertainty contributions listing. Required for some ISO/IEC 17025 and defense laboratory applications. Available on request.
Supplied at time of manufacture or as standalone. Standard or custom setpoints. AMS 2750F and pharma validation documentation.