Calibration Services · NIST-Traceable

NIST-Traceable
Calibration
Certificates

Individual calibration certificates with expanded measurement uncertainty statements (k=2) per ISO/IEC 17025 framework — for RTD sensors and thermocouple assemblies at customer-specified temperature setpoints. Supplied at time of manufacture or as a standalone calibration service.

Calibration Process Overview
1
Reference Standard Selection
NIST-traceable reference — PT25 or Type S/R noble metal TC. Calibration chain documented.
2
Setpoint Conditioning
Bath or furnace stabilized at each customer setpoint. Stability criterion met before measurement.
3
Data Acquisition
Reference and DUT measured simultaneously. Multiple readings per setpoint for statistical averaging.
4
Uncertainty Analysis
Combined uncertainty calculated per GUM (JCGM 100:2008). Expanded uncertainty at k=2 (95% confidence).
5
Certificate Issuance
Individual cal cert with serial number, setpoint data, correction values, and uncertainty statement. Signed by Quality.
Calibration Services

What We Calibrate

🌡️
RTD / PT100 Calibration

NIST-traceable resistance vs. temperature calibration for PT100, PT500, and PT1000 RTD assemblies. Data reported as resistance (Ω) at each setpoint, with deviation from IEC 60751 nominal. Useful for pharmaceutical validation, process qualification, and high-accuracy measurement systems.

Class AA, A, B element types
2-wire, 3-wire, or 4-wire measurement
Setpoints from –80°C to +550°C
Resistance data at each point (Ω)
Deviation from IEC 60751 nominal
Expanded uncertainty at k=2
⚡
Thermocouple Calibration

EMF vs. temperature calibration for base metal (K, J, T, N, E) and noble metal (R, S, B) thermocouple assemblies per IEC 60584-1. Data reported as EMF (mV) at each setpoint, with deviation from IEC 60584 reference function. Required for AMS 2750F SAT sensor qualification and process validation.

Types K, J, T, N, E, R, S, B
MIMS and conventional construction
Setpoints from –50°C to +1300°C
EMF data (mV) at each setpoint
Deviation from IEC 60584 reference
AMS 2750F SAT documentation
📋
Validation & Qualification Sets

Sets of matched sensors calibrated together for Temperature Uniformity Survey (TUS), process qualification, and validation protocols. Each sensor receives an individual certificate with a common set serial number for batch traceability. Lyophilizer mapping, autoclave validation, and furnace TUS applications.

Sets of 2–30 sensors
Individual serial number per sensor
Common batch/set reference
Matched tolerance for uniformity studies
Lyophilizer, autoclave, furnace TUS
IQ/OQ documentation support
Applications

Who Requires NIST-Traceable Calibration?

✈️
Aerospace Heat Treating

AMS 2750F requires calibrated TCs for SAT (System Accuracy Test) sensors. NADCAP suppliers require individual cal certs with current calibration dates and traceability chain.

AMS 2750F · NADCAP
💊
Pharmaceutical Manufacturing

21 CFR Part 11 and EU GMP require documented calibration traceability for all process monitoring instruments used in product manufacturing and storage.

21 CFR · EU GMP Annex 1
🧊
Autoclave & Lyophilizer Validation

Temperature uniformity mapping and validation protocols require NIST-traceable cal certs with uncertainty statements for each sensor used in IQ/OQ/PQ testing.

USP <1211> · EN 285 · HTM 2010
🏭
Industrial Process Control

ISO 9001:2015 requires measurement equipment to be calibrated at specified intervals against measurement standards traceable to national or international standards.

ISO 9001:2015 Cl. 7.1.5.2
🔬
R&D and Test Laboratories

Research publications require measurement traceability to SI. Test reports must include calibration data and uncertainty. ISO/IEC 17025 accredited lab requirements.

ISO/IEC 17025 · GUM
⚡
Power Generation

Turbine and boiler temperature monitoring for NERC compliance and insurance requirements. Calibrated sensors with documented traceability for critical protection systems.

NERC · API 670
🛢️
Oil & Gas Process Safety

Safety Instrumented Systems (SIS) per IEC 61511 require documented instrument calibration and proof testing intervals for sensors in safety instrumented functions (SIF).

IEC 61511 · API 670
🍔
Food Processing & HACCP

HACCP critical control point temperatures require calibrated sensors with documented traceability. FDA inspection may require calibration records for CCP monitoring instruments.

HACCP · FDA 21 CFR
Calibration Process

How Thermometrics Calibrates

01
Reference Standard

Calibrated against NIST-traceable reference — PT25 capsule SPRT or noble metal TC with current calibration certificate. Traceability chain maintained and documented.

02
Thermal Stabilization

Temperature bath or calibration furnace stabilized at setpoint. Stability criterion: ≤0.05°C drift over 5 minutes before measurement begins.

03
Simultaneous Measurement

Reference and DUT measured by calibrated readout instruments (Fluke 1594A or equivalent). Minimum 5 readings per setpoint recorded and averaged.

04
Uncertainty Budget

Combined standard uncertainty calculated per GUM (JCGM 100:2008). Includes reference uncertainty, repeatability, readout resolution, and thermal gradient contributions.

05
Certificate Issue

Individual PDF certificate with unique serial number, setpoint data table, correction values, expanded uncertainty (k=2), and QA signature. Issued per AS9100D QMS.

Documentation Deliverables

What's on the Calibration Certificate

STANDARD
Sensor Identification

Customer PO number, Thermometrics order number, sensor model, unique serial number, and date of calibration.

STANDARD
Setpoint Data Table

Temperature setpoints in °C, measured resistance (RTD in Ω) or EMF (TC in mV), deviation from IEC nominal, and correction factor at each point.

STANDARD
Expanded Uncertainty

Expanded measurement uncertainty (U) at 95% confidence level (coverage factor k=2) per GUM. Stated at each setpoint or as a single worst-case value.

STANDARD
Traceability Statement

Statement confirming traceability to NIST through an unbroken chain of comparisons with stated uncertainties. Reference standard ID and calibration date included.

STANDARD
QA Signature & Stamp

Signed by Thermometrics Quality Assurance. AS9100D document control number. Certificate revision control.

ON REQUEST
Uncertainty Budget Worksheet

Detailed Type A and Type B uncertainty contributions listing. Required for some ISO/IEC 17025 and defense laboratory applications. Available on request.

Request NIST-Traceable Calibration

Supplied at time of manufacture or as standalone. Standard or custom setpoints. AMS 2750F and pharma validation documentation.