Thermocouple EMF Tables
Complete millivolt (mV) output vs. temperature reference data for all standard thermocouple types per IEC 60584 — Type K, J, T, N, E, R, S, and B. Reference junction at 0°C. Includes cold junction compensation explanation, sensitivity data, and live interactive EMF calculator.
| Type | Alloy (+/–) | Range | Sensitivity at 25°C | Class 1 Tol. at 100°C | Atmosphere | Key Strength |
|---|---|---|---|---|---|---|
| K | Chromel / Alumel | –200 to +1260°C | ~41 µV/°C | ±1.5°C | Oxidizing, inert | Most widely used, broad range |
| J | Iron / Constantan | –40 to +750°C | ~51 µV/°C | ±1.5°C | Reducing, vacuum | High sensitivity, reducing atmospheres |
| T | Copper / Constantan | –270 to +400°C | ~40 µV/°C | ±0.5°C | Oxidizing, reducing | Best sub-zero, moisture resistant |
| N | Nicrosil / Nisil | –270 to +1300°C | ~26 µV/°C | ±1.5°C | Oxidizing, inert | Long-term stability, low drift vs. K |
| E | Chromel / Constantan | –40 to +900°C | ~62 µV/°C | ±1.5°C | Oxidizing, inert | Highest sensitivity of base metal types |
| R | Pt13%Rh / Pt | –50 to +1768°C | ~10 µV/°C | ±1.0°C | Oxidizing, inert | High accuracy, SAT reference |
| S | Pt10%Rh / Pt | –50 to +1768°C | ~10 µV/°C | ±1.0°C | Oxidizing, inert | High accuracy, primary calibration standard |
| B | Pt30%Rh / Pt6%Rh | +600 to +1820°C | ~6 µV/°C | ±0.5% above 1100°C | Oxidizing, inert | Ultra-high temp, near-zero EMF at RT |
Type K — Chromel / Alumel
Type K is the most widely used thermocouple type. EMF values in millivolts, reference junction at 0°C. Sensitivity peaks near 600°C (~43 µV/°C). Subject to "K-effect" reversible drift between 200–600°C from Ni-Cr ordering. Above 500°C in air, Inconel sheath recommended.
| Temp °C | 0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| –200 | –5.891 | –5.730 | –5.550 | –5.354 | –5.141 | –4.913 | –4.669 | –4.411 | –4.138 | –3.852 |
| –100 | –3.554 | –3.243 | –2.920 | –2.587 | –2.243 | –1.889 | –1.527 | –1.156 | –0.778 | –0.392 |
| 0 | 0.000 | 0.397 | 0.798 | 1.203 | 1.612 | 2.023 | 2.436 | 2.851 | 3.267 | 3.682 |
| 100 | 4.096 | 4.509 | 4.920 | 5.328 | 5.735 | 6.138 | 6.540 | 6.941 | 7.340 | 7.739 |
| 200 | 8.138 | 8.539 | 8.940 | 9.343 | 9.747 | 10.153 | 10.561 | 10.971 | 11.382 | 11.795 |
| 300 | 12.209 | 12.624 | 13.040 | 13.457 | 13.874 | 14.293 | 14.713 | 15.133 | 15.554 | 15.975 |
| 400 | 16.397 | 16.820 | 17.243 | 17.667 | 18.091 | 18.516 | 18.941 | 19.366 | 19.792 | 20.218 |
| 500 | 20.644 | 21.071 | 21.497 | 21.924 | 22.350 | 22.776 | 23.203 | 23.629 | 24.055 | 24.480 |
| 600 | 24.905 | 25.330 | 25.755 | 26.179 | 26.602 | 27.025 | 27.447 | 27.869 | 28.289 | 28.710 |
| 700 | 29.129 | 29.548 | 29.965 | 30.382 | 30.798 | 31.213 | 31.628 | 32.041 | 32.453 | 32.865 |
| 800 | 33.275 | 33.685 | 34.093 | 34.501 | 34.908 | 35.313 | 35.718 | 36.121 | 36.524 | 36.925 |
| 900 | 37.326 | 37.725 | 38.124 | 38.522 | 38.918 | 39.314 | 39.708 | 40.101 | 40.494 | 40.885 |
| 1000 | 41.276 | 41.665 | 42.053 | 42.440 | 42.826 | 43.211 | 43.595 | 43.978 | 44.359 | 44.740 |
| 1100 | 45.119 | 45.497 | 45.873 | 46.249 | 46.623 | 46.995 | 47.367 | 47.737 | 48.105 | 48.473 |
| 1200 | 48.838 | 49.202 | 49.565 | 49.926 | 50.286 | 50.644 | 51.000 | 51.355 | — | — |
Type J — Iron / Constantan
Type J has the highest sensitivity of the common base metal types below 500°C and performs well in reducing and vacuum atmospheres. Iron positive leg oxidizes rapidly above 550°C in air — do not use above 500°C in oxidizing atmosphere without Inconel sheath. Not recommended for use below –40°C.
| Temp °C | 0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| –40 | –1.961 | –1.482 | –0.995 | –0.501 | –0.000 | — | — | — | — | — |
| 0 | 0.000 | 0.507 | 1.019 | 1.537 | 2.059 | 2.585 | 3.116 | 3.650 | 4.187 | 4.726 |
| 100 | 5.269 | 5.814 | 6.360 | 6.909 | 7.459 | 8.010 | 8.562 | 9.115 | 9.669 | 10.224 |
| 200 | 10.779 | 11.334 | 11.889 | 12.445 | 13.000 | 13.555 | 14.110 | 14.665 | 15.219 | 15.773 |
| 300 | 16.327 | 16.881 | 17.434 | 17.986 | 18.538 | 19.090 | 19.642 | 20.194 | 20.745 | 21.297 |
| 400 | 21.848 | 22.400 | 22.952 | 23.504 | 24.057 | 24.610 | 25.164 | 25.720 | 26.276 | 26.834 |
| 500 | 27.393 | 27.953 | 28.516 | 29.080 | 29.647 | 30.216 | 30.788 | 31.362 | 31.939 | 32.519 |
| 600 | 33.102 | 33.689 | 34.279 | 34.873 | 35.470 | 36.071 | 36.675 | 37.284 | 37.896 | 38.512 |
| 700 | 39.132 | 39.755 | 40.382 | 41.012 | 41.645 | 42.281 | 42.919 | 43.559 | 44.203 | 44.848 |
Type T — Copper / Constantan
Type T is the preferred type for sub-zero and cryogenic temperature measurement and for any application involving moisture or condensation. Copper positive leg provides high thermal conductivity — minimise stem conduction error in high-accuracy applications. Upper limit +400°C (copper oxidises above 370°C in air).
| Temp °C | 0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| –200 | –5.603 | –5.439 | –5.261 | –5.069 | –4.865 | –4.648 | –4.419 | –4.177 | –3.923 | –3.657 |
| –100 | –3.379 | –3.089 | –2.788 | –2.476 | –2.153 | –1.819 | –1.475 | –1.121 | –0.757 | –0.383 |
| 0 | 0.000 | 0.391 | 0.790 | 1.196 | 1.612 | 2.036 | 2.468 | 2.909 | 3.358 | 3.814 |
| 100 | 4.279 | 4.750 | 5.228 | 5.714 | 6.206 | 6.704 | 7.209 | 7.720 | 8.237 | 8.759 |
| 200 | 9.288 | 9.822 | 10.362 | 10.907 | 11.458 | 12.013 | 12.574 | 13.139 | 13.709 | 14.283 |
| 300 | 14.862 | 15.445 | 16.032 | 16.624 | 17.219 | 17.819 | 18.422 | 19.030 | 19.641 | 20.255 |
| 400 | 20.872 | — | — | — | — | — | — | — | — | — |
Type N — Nicrosil / Nisil
Type N was engineered specifically to overcome the limitations of Type K — particularly short-range ordering instability between 200–600°C. It demonstrates 3 to 10 times less drift than Type K under equivalent conditions above 300°C. Preferred for new industrial furnace installations requiring long calibration intervals. Sensitivity is lower than Type K (~26 µV/°C vs. 41 µV/°C).
| Temp °C | 0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| –200 | –3.990 | –3.884 | –3.766 | –3.634 | –3.491 | –3.336 | –3.171 | –2.994 | –2.808 | –2.612 |
| –100 | –2.407 | –2.193 | –1.972 | –1.744 | –1.509 | –1.269 | –1.023 | –0.772 | –0.518 | –0.260 |
| 0 | 0.000 | 0.261 | 0.525 | 0.793 | 1.065 | 1.340 | 1.619 | 1.902 | 2.189 | 2.480 |
| 100 | 2.774 | 3.072 | 3.374 | 3.680 | 3.989 | 4.302 | 4.618 | 4.937 | 5.259 | 5.585 |
| 200 | 5.913 | 6.245 | 6.579 | 6.916 | 7.255 | 7.597 | 7.941 | 8.288 | 8.637 | 8.988 |
| 300 | 9.341 | 9.696 | 10.054 | 10.413 | 10.774 | 11.136 | 11.501 | 11.867 | 12.234 | 12.603 |
| 400 | 12.974 | 13.346 | 13.719 | 14.094 | 14.469 | 14.846 | 15.225 | 15.604 | 15.984 | 16.366 |
| 500 | 16.748 | 17.131 | 17.515 | 17.900 | 18.286 | 18.672 | 19.059 | 19.447 | 19.835 | 20.224 |
| 600 | 20.613 | 21.003 | 21.393 | 21.784 | 22.175 | 22.566 | 22.958 | 23.350 | 23.742 | 24.134 |
| 900 | 30.934 | 31.332 | 31.729 | 32.127 | 32.524 | 32.921 | 33.318 | 33.715 | 34.112 | 34.508 |
| 1200 | 42.164 | 42.556 | 42.947 | 43.337 | 43.726 | 44.115 | 44.503 | 44.891 | — | — |
Type E — Chromel / Constantan
Type E produces the highest EMF output of all standardized thermocouple types (~62 µV/°C near room temperature), making it ideal for high-resolution measurement and differential temperature sensing. Performs well at sub-zero temperatures. Susceptible to preferential oxidation of the chromium in the positive leg above 900°C in low-oxygen environments (green rot).
| Temp °C | 0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| –100 | –5.237 | –4.777 | –4.302 | –3.811 | –3.306 | –2.787 | –2.255 | –1.709 | –1.152 | –0.582 |
| 0 | 0.000 | 0.591 | 1.192 | 1.801 | 2.420 | 3.048 | 3.685 | 4.330 | 4.985 | 5.648 |
| 100 | 6.319 | 6.998 | 7.685 | 8.379 | 9.081 | 9.789 | 10.503 | 11.224 | 11.951 | 12.684 |
| 200 | 13.421 | 14.164 | 14.912 | 15.664 | 16.420 | 17.181 | 17.945 | 18.713 | 19.484 | 20.259 |
| 300 | 21.036 | 21.817 | 22.600 | 23.386 | 24.174 | 24.964 | 25.757 | 26.552 | 27.348 | 28.146 |
| 400 | 28.946 | 29.747 | 30.550 | 31.354 | 32.159 | 32.965 | 33.772 | 34.579 | 35.387 | 36.196 |
| 500 | 37.005 | 37.815 | 38.624 | 39.434 | 40.243 | 41.053 | 41.862 | 42.671 | 43.479 | 44.286 |
| 600 | 45.093 | 45.900 | 46.705 | 47.509 | 48.313 | 49.116 | 49.917 | 50.718 | 51.517 | 52.315 |
Type R — Pt13%Rh / Platinum
Noble metal type. Type R provides high accuracy and long-term stability in oxidizing and inert atmospheres to 1768°C. Required for System Accuracy Tests (SAT) per AMS 2750F in precision furnace classes. EMF values are significantly lower than base metal types — high-quality instrument with good resolution required. Platinum contaminated by base metals (particularly iron) will drift irreversibly.
| Temp °C | 0 | +100 | +200 | +300 | +400 | +500 | +600 | +700 | +800 | +900 |
|---|---|---|---|---|---|---|---|---|---|---|
| 0 | 0.000 | 0.647 | 1.469 | 2.401 | 3.408 | 4.471 | 5.583 | 6.743 | 7.950 | 9.203 |
| 1000 | 10.503 | 11.846 | 13.228 | 14.629 | 16.040 | 17.451 | 18.849 | 20.222 | 21.570 | 22.887 |
| 1600 | 21.103 | — | — | — | — | — | — | — | — | — |
Type S — Pt10%Rh / Platinum
The Type S thermocouple was historically the primary international temperature standard for the range 630.74°C to 1064.43°C under IPTS-68. It remains the reference type for high-precision calibration and is widely used as a SAT reference sensor per AMS 2750F. Type S has slightly lower EMF output than Type R at equivalent temperatures. Same contamination sensitivity as Type R — protect from base metal vapors.
| Temp °C | 0 | +100 | +200 | +300 | +400 | +500 | +600 | +700 | +800 | +900 |
|---|---|---|---|---|---|---|---|---|---|---|
| 0 | 0.000 | 0.646 | 1.441 | 2.323 | 3.260 | 4.233 | 5.239 | 6.275 | 7.345 | 8.449 |
| 1000 | 9.587 | 10.757 | 11.951 | 13.159 | 14.373 | 15.582 | 16.777 | 17.947 | 19.090 | 20.200 |
| 1600 | 18.503 | — | — | — | — | — | — | — | — | — |
Type B — Pt30%Rh / Pt6%Rh
Type B is used exclusively for very high temperature applications above 600°C in oxidizing or inert atmospheres. Unique characteristic: near-zero EMF at room temperature (output at 0°C ≈ 0 µV, at 42°C ≈ 0 µV) — making it insensitive to cold junction temperature variation below ~100°C. This eliminates the need for cold junction compensation for most installations, but also means it cannot measure below ~600°C with useful accuracy. Sensitivity is the lowest of all standardized types (~6 µV/°C).
| Temp °C | 0 | +100 | +200 | +300 | +400 | +500 | +600 | +700 | +800 | +900 |
|---|---|---|---|---|---|---|---|---|---|---|
| 0 | 0.000 | 0.033 | 0.178 | 0.431 | 0.787 | 1.242 | 1.792 | 2.431 | 3.154 | 3.957 |
| 1000 | 4.834 | 5.780 | 6.786 | 7.848 | 8.956 | 10.099 | 11.263 | 12.433 | 13.591 | 14.720 |
| 1600 | 12.426 | 13.099 | 13.769 | — | — | — | — | — | — | — |
Cold Junction Compensation (CJC)
All thermocouple EMF tables are referenced to 0°C (the ice point). In practice, the cold junction (instrument terminals, transmitter, or isothermal block) is at ambient temperature — not 0°C. Cold junction compensation (CJC) accounts for the difference between the actual cold junction temperature and 0°C.
Most modern thermocouple transmitters and instruments include an on-board thermistor or RTD to measure the cold junction temperature, and perform this compensation automatically using the polynomial coefficients from IEC 60584. In a DCS or data acquisition system, the CJC sensor is typically located at the isothermal terminal block where extension wire is joined to copper instrument wiring.
Critical installation requirement: thermocouple extension wire must use the correct type-matched alloy from the TC hot junction to the instrument terminals. Connecting copper wire or mismatched extension cable at any intermediate point introduces a parasitic thermocouple junction at that splice, generating an unknown temperature-dependent EMF error that cannot be compensated by the instrument's CJC circuit.