Technical Reference · IEC 60584

Thermocouple EMF Tables

Complete millivolt (mV) output vs. temperature reference data for all standard thermocouple types per IEC 60584 — Type K, J, T, N, E, R, S, and B. Reference junction at 0°C. Includes cold junction compensation explanation, sensitivity data, and live interactive EMF calculator.

IEC 60584:2013Types K J T N E R S B–270°C to +1820°CCold Junction CompensationPrint Friendly
Quick Reference
Thermocouple Type Comparison — IEC 60584
Thermocouple Type Comparison — Key Properties per IEC 60584
TypeAlloy (+/–)RangeSensitivity at 25°CClass 1 Tol. at 100°CAtmosphereKey Strength
KChromel / Alumel–200 to +1260°C~41 µV/°C±1.5°COxidizing, inertMost widely used, broad range
JIron / Constantan–40 to +750°C~51 µV/°C±1.5°CReducing, vacuumHigh sensitivity, reducing atmospheres
TCopper / Constantan–270 to +400°C~40 µV/°C±0.5°COxidizing, reducingBest sub-zero, moisture resistant
NNicrosil / Nisil–270 to +1300°C~26 µV/°C±1.5°COxidizing, inertLong-term stability, low drift vs. K
EChromel / Constantan–40 to +900°C~62 µV/°C±1.5°COxidizing, inertHighest sensitivity of base metal types
RPt13%Rh / Pt–50 to +1768°C~10 µV/°C±1.0°COxidizing, inertHigh accuracy, SAT reference
SPt10%Rh / Pt–50 to +1768°C~10 µV/°C±1.0°COxidizing, inertHigh accuracy, primary calibration standard
BPt30%Rh / Pt6%Rh+600 to +1820°C~6 µV/°C±0.5% above 1100°COxidizing, inertUltra-high temp, near-zero EMF at RT

Type K — Chromel / Alumel

IEC 60584-1 · Positive: Chromel (Ni-10%Cr) · Negative: Alumel (Ni-2%Al-2%Mn-1%Si)

Type K is the most widely used thermocouple type. EMF values in millivolts, reference junction at 0°C. Sensitivity peaks near 600°C (~43 µV/°C). Subject to "K-effect" reversible drift between 200–600°C from Ni-Cr ordering. Above 500°C in air, Inconel sheath recommended.

Type K EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+10+20+30+40+50+60+70+80+90
–200–5.891–5.730–5.550–5.354–5.141–4.913–4.669–4.411–4.138–3.852
–100–3.554–3.243–2.920–2.587–2.243–1.889–1.527–1.156–0.778–0.392
00.0000.3970.7981.2031.6122.0232.4362.8513.2673.682
1004.0964.5094.9205.3285.7356.1386.5406.9417.3407.739
2008.1388.5398.9409.3439.74710.15310.56110.97111.38211.795
30012.20912.62413.04013.45713.87414.29314.71315.13315.55415.975
40016.39716.82017.24317.66718.09118.51618.94119.36619.79220.218
50020.64421.07121.49721.92422.35022.77623.20323.62924.05524.480
60024.90525.33025.75526.17926.60227.02527.44727.86928.28928.710
70029.12929.54829.96530.38230.79831.21331.62832.04132.45332.865
80033.27533.68534.09334.50134.90835.31335.71836.12136.52436.925
90037.32637.72538.12438.52238.91839.31439.70840.10140.49440.885
100041.27641.66542.05342.44042.82643.21143.59543.97844.35944.740
110045.11945.49745.87346.24946.62346.99547.36747.73748.10548.473
120048.83849.20249.56549.92650.28650.64451.00051.355——

Type J — Iron / Constantan

IEC 60584-1 · Positive: Iron (Fe) · Negative: Constantan (Cu-45%Ni)

Type J has the highest sensitivity of the common base metal types below 500°C and performs well in reducing and vacuum atmospheres. Iron positive leg oxidizes rapidly above 550°C in air — do not use above 500°C in oxidizing atmosphere without Inconel sheath. Not recommended for use below –40°C.

Type J EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+10+20+30+40+50+60+70+80+90
–40–1.961–1.482–0.995–0.501–0.000—————
00.0000.5071.0191.5372.0592.5853.1163.6504.1874.726
1005.2695.8146.3606.9097.4598.0108.5629.1159.66910.224
20010.77911.33411.88912.44513.00013.55514.11014.66515.21915.773
30016.32716.88117.43417.98618.53819.09019.64220.19420.74521.297
40021.84822.40022.95223.50424.05724.61025.16425.72026.27626.834
50027.39327.95328.51629.08029.64730.21630.78831.36231.93932.519
60033.10233.68934.27934.87335.47036.07136.67537.28437.89638.512
70039.13239.75540.38241.01241.64542.28142.91943.55944.20344.848

Type T — Copper / Constantan

IEC 60584-1 · Positive: Copper (Cu) · Negative: Constantan (Cu-45%Ni)

Type T is the preferred type for sub-zero and cryogenic temperature measurement and for any application involving moisture or condensation. Copper positive leg provides high thermal conductivity — minimise stem conduction error in high-accuracy applications. Upper limit +400°C (copper oxidises above 370°C in air).

Type T EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+10+20+30+40+50+60+70+80+90
–200–5.603–5.439–5.261–5.069–4.865–4.648–4.419–4.177–3.923–3.657
–100–3.379–3.089–2.788–2.476–2.153–1.819–1.475–1.121–0.757–0.383
00.0000.3910.7901.1961.6122.0362.4682.9093.3583.814
1004.2794.7505.2285.7146.2066.7047.2097.7208.2378.759
2009.2889.82210.36210.90711.45812.01312.57413.13913.70914.283
30014.86215.44516.03216.62417.21917.81918.42219.03019.64120.255
40020.872—————————

Type N — Nicrosil / Nisil

IEC 60584-1 · Positive: Nicrosil (Ni-14.2%Cr-1.4%Si) · Negative: Nisil (Ni-4.4%Si-0.1%Mg)

Type N was engineered specifically to overcome the limitations of Type K — particularly short-range ordering instability between 200–600°C. It demonstrates 3 to 10 times less drift than Type K under equivalent conditions above 300°C. Preferred for new industrial furnace installations requiring long calibration intervals. Sensitivity is lower than Type K (~26 µV/°C vs. 41 µV/°C).

Type N EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+10+20+30+40+50+60+70+80+90
–200–3.990–3.884–3.766–3.634–3.491–3.336–3.171–2.994–2.808–2.612
–100–2.407–2.193–1.972–1.744–1.509–1.269–1.023–0.772–0.518–0.260
00.0000.2610.5250.7931.0651.3401.6191.9022.1892.480
1002.7743.0723.3743.6803.9894.3024.6184.9375.2595.585
2005.9136.2456.5796.9167.2557.5977.9418.2888.6378.988
3009.3419.69610.05410.41310.77411.13611.50111.86712.23412.603
40012.97413.34613.71914.09414.46914.84615.22515.60415.98416.366
50016.74817.13117.51517.90018.28618.67219.05919.44719.83520.224
60020.61321.00321.39321.78422.17522.56622.95823.35023.74224.134
90030.93431.33231.72932.12732.52432.92133.31833.71534.11234.508
120042.16442.55642.94743.33743.72644.11544.50344.891——

Type E — Chromel / Constantan

IEC 60584-1 · Positive: Chromel (Ni-10%Cr) · Negative: Constantan (Cu-45%Ni)

Type E produces the highest EMF output of all standardized thermocouple types (~62 µV/°C near room temperature), making it ideal for high-resolution measurement and differential temperature sensing. Performs well at sub-zero temperatures. Susceptible to preferential oxidation of the chromium in the positive leg above 900°C in low-oxygen environments (green rot).

Type E EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+10+20+30+40+50+60+70+80+90
–100–5.237–4.777–4.302–3.811–3.306–2.787–2.255–1.709–1.152–0.582
00.0000.5911.1921.8012.4203.0483.6854.3304.9855.648
1006.3196.9987.6858.3799.0819.78910.50311.22411.95112.684
20013.42114.16414.91215.66416.42017.18117.94518.71319.48420.259
30021.03621.81722.60023.38624.17424.96425.75726.55227.34828.146
40028.94629.74730.55031.35432.15932.96533.77234.57935.38736.196
50037.00537.81538.62439.43440.24341.05341.86242.67143.47944.286
60045.09345.90046.70547.50948.31349.11649.91750.71851.51752.315

Type R — Pt13%Rh / Platinum

IEC 60584-1 · Positive: Platinum-13% Rhodium (Pt-13%Rh) · Negative: Platinum (Pt)

Noble metal type. Type R provides high accuracy and long-term stability in oxidizing and inert atmospheres to 1768°C. Required for System Accuracy Tests (SAT) per AMS 2750F in precision furnace classes. EMF values are significantly lower than base metal types — high-quality instrument with good resolution required. Platinum contaminated by base metals (particularly iron) will drift irreversibly.

Type R EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+100+200+300+400+500+600+700+800+900
00.0000.6471.4692.4013.4084.4715.5836.7437.9509.203
100010.50311.84613.22814.62916.04017.45118.84920.22221.57022.887
160021.103—————————

Type S — Pt10%Rh / Platinum

IEC 60584-1 · Positive: Platinum-10% Rhodium (Pt-10%Rh) · Negative: Platinum (Pt)

The Type S thermocouple was historically the primary international temperature standard for the range 630.74°C to 1064.43°C under IPTS-68. It remains the reference type for high-precision calibration and is widely used as a SAT reference sensor per AMS 2750F. Type S has slightly lower EMF output than Type R at equivalent temperatures. Same contamination sensitivity as Type R — protect from base metal vapors.

Type S EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+100+200+300+400+500+600+700+800+900
00.0000.6461.4412.3233.2604.2335.2396.2757.3458.449
10009.58710.75711.95113.15914.37315.58216.77717.94719.09020.200
160018.503—————————

Type B — Pt30%Rh / Pt6%Rh

IEC 60584-1 · Positive: Platinum-30% Rhodium (Pt-30%Rh) · Negative: Platinum-6% Rhodium (Pt-6%Rh)

Type B is used exclusively for very high temperature applications above 600°C in oxidizing or inert atmospheres. Unique characteristic: near-zero EMF at room temperature (output at 0°C ≈ 0 µV, at 42°C ≈ 0 µV) — making it insensitive to cold junction temperature variation below ~100°C. This eliminates the need for cold junction compensation for most installations, but also means it cannot measure below ~600°C with useful accuracy. Sensitivity is the lowest of all standardized types (~6 µV/°C).

Type B EMF (mV) — IEC 60584-1 — Reference junction 0°C
Temp °C0+100+200+300+400+500+600+700+800+900
00.0000.0330.1780.4310.7871.2421.7922.4313.1543.957
10004.8345.7806.7867.8488.95610.09911.26312.43313.59114.720
160012.42613.09913.769———————

Cold Junction Compensation (CJC)

All thermocouple EMF tables are referenced to 0°C (the ice point). In practice, the cold junction (instrument terminals, transmitter, or isothermal block) is at ambient temperature — not 0°C. Cold junction compensation (CJC) accounts for the difference between the actual cold junction temperature and 0°C.

V_measured = V(T_hot, 0°C) – V(T_cold, 0°C)
T_hot = EMF⁻¹( V_measured + V(T_cold, 0°C) )
/* V(T,0) = EMF at temperature T with reference at 0°C, from IEC 60584 tables */

Most modern thermocouple transmitters and instruments include an on-board thermistor or RTD to measure the cold junction temperature, and perform this compensation automatically using the polynomial coefficients from IEC 60584. In a DCS or data acquisition system, the CJC sensor is typically located at the isothermal terminal block where extension wire is joined to copper instrument wiring.

Critical installation requirement: thermocouple extension wire must use the correct type-matched alloy from the TC hot junction to the instrument terminals. Connecting copper wire or mismatched extension cable at any intermediate point introduces a parasitic thermocouple junction at that splice, generating an unknown temperature-dependent EMF error that cannot be compensated by the instrument's CJC circuit.