MIMS and conventional thermocouple assemblies in Types K, J, T, N, E, R, S, and B — manufactured to IEC 60584 Class 1 and Class 2 tolerance in Inconel 600, Inconel 625, Hastelloy C-276, and 316 SS sheaths. Any length, any termination. AS9100D quality documentation standard.
Every standard thermocouple type manufactured in-house to IEC 60584 tolerance — from the general-purpose Type K to noble metal Type R/S SAT reference sensors. Custom alloy combinations and non-standard types on request.
The world's most-used thermocouple. Wide temperature range, good linearity (~41 µV/°C), and universal instrument support. Subject to short-range ordering (K-effect) in 200–650°C service — Type N is the stability upgrade for long-term furnace applications. Positive leg color: Green (IEC), Yellow (ANSI).
Higher sensitivity than Type K (~52 µV/°C). Preferred for reducing and vacuum atmospheres where iron is stable but chromium-containing alloys would preferentially oxidize. Iron leg corrodes rapidly above 550°C in air and in wet environments. Limit continuous use in oxidizing atmosphere to ~500°C. Iron leg is magnetic — small Curie point perturbation at 770°C.
The preferred thermocouple for cryogenic, refrigeration, pharmaceutical cold chain, and food processing applications. Stable and well-characterized at sub-zero temperatures. Both alloy legs resist corrosion in moist environments. Copper positive leg limits maximum temperature to +400°C. IEC Class 1 tolerance of ±0.5°C is the tightest fixed tolerance of any standard base-metal type.
Engineered to correct the drift mechanisms in Type K. Silicon additions to both legs stabilize the alloy microstructure and suppress the ordering transformation responsible for K-effect drift. Demonstrates 3–10× less drift than Type K over equivalent service life. Preferred for new furnace installations above 500°C and for AMS 2750F heat treating where calibration interval reduction is desired.
Type E has the highest Seebeck coefficient (~68 µV/°C at 0°C) of all IEC-standardized thermocouple types — approximately 65% higher than Type K at the same temperature. This makes Type E best suited for detecting small temperature differences, differential temperature measurement, and applications where measurement resolution is critical. Non-magnetic, which is an advantage in some EMI environments. Limited to +1000°C maximum.
Noble metal platinum-rhodium thermocouples offering the highest accuracy and stability of any IEC-standardized type in oxidizing atmospheres above 600°C. IEC Class 1 tolerance of ±1.0°C at 1000°C. Required for AMS 2750F SAT (System Accuracy Test) sensors and NIST-traceable reference applications. NIST-traceable calibration certificate is standard at Thermometrics for all noble metal shipments.
Type B is the only IEC-standardized thermocouple type suitable for continuous service above 1600°C in oxidizing or neutral atmospheres. Both legs contain rhodium, producing near-zero EMF at room temperature — this means Type B cannot distinguish temperatures below approximately 50°C and must not be used below +600°C. Very low sensitivity (~6 µV/°C) requires careful instrument selection. Superior contamination resistance vs. Types R and S due to absence of pure platinum leg.
Type C (W5%Re/W26%Re) for vacuum/inert atmospheres to 2315°C. Non-standard alloy pairs for specific industrial requirements. Dual-element redundant assemblies. Multi-point sensor arrays. Assembly-level test jigs. Contact engineering to discuss your specific high-temperature or special-alloy requirement.
All Thermometrics thermocouples are available in MIMS (Mineral Insulated Metal Sheath) or conventional (ceramic/fiberglass insulated) construction. MIMS is the standard for industrial, aerospace, and pressure service applications.
MIMS thermocouples are manufactured by drawing a seamless metal sheath tube containing thermocouple wire conductors embedded in compacted magnesium oxide (MgO) powder insulation. The drawing process densifies the MgO to >85% theoretical density, locking the conductors in position and providing excellent vibration resistance, pressure integrity, and high electrical insulation resistance (≥100 MΩ at 500 VDC).
Conventional thermocouples consist of bare thermocouple wire conductors insulated with ceramic tubes, fiberglass braiding, or PTFE insulation. No outer metal sheath (or a protection tube added separately). Lower cost for basic applications but limited in vibration resistance, minimum bend radius, and pressure compatibility.
Sheath material selection is determined by process temperature, atmosphere chemistry, and mechanical requirements. Thermometrics stocks standard MIMS sheath materials and can source special alloys on request with mill certification.
| Alloy | UNS | Max Continuous | Key Properties | Primary Application |
|---|---|---|---|---|
| 304 SS | S30400 | 900°C (oxidizing) | General purpose, cost-effective, non-magnetic option available | General industrial, HVAC, low-pressure service |
| 316 SS | S31600 | 900°C | Improved chloride and pitting corrosion resistance vs. 304 | Standard process, mild corrosive environments |
| 316L SS | S31603 | 900°C | Low carbon — resists weld sensitization, FDA/3-A sanitary | Pharmaceutical, food processing, CIP/SIP |
| Inconel 600 | N06600 | 1175°C | Outstanding high-temperature oxidation resistance, Ni-Cr alloy | Furnace, heat treating, aerospace, high-temp TC |
| Inconel 625 | N06625 | 980°C | H₂S and seawater resistant, high strength, NACE MR0175 compatible | Sour service, subsea, O&G, marine |
| Hastelloy C-276 | N10276 | 1040°C | Exceptional resistance to oxidizing and reducing corrosion, acids, chlorides | Chemical processing, chloride media, acid service |
| Incoloy 800H/HT | N08810/N08811 | 1100°C | High-temperature creep strength and oxidation resistance | High-temperature furnace, petrochemical reformers |
| Custom Alloy | Per request | Per alloy | Titanium, Duplex, SMO 254, platinum-clad — on request | Contact engineering for non-standard sheath requirements |
For MIMS thermocouples, three junction styles are available — each with different response time, electrical isolation, and application suitability characteristics.
The welded junction is in direct contact with the sheath tip. Fastest response time of the three MIMS styles — typically 25–50% faster than ungrounded. The electrical connection to the sheath means ground loops can occur in multi-point systems with common ground instrumentation. Use when response time is critical and electrical isolation is not required.
The welded junction is electrically isolated from the sheath by a thin layer of compacted MgO powder between the junction and the sheath tip. Electrically isolates the measurement circuit from the process and sheath — eliminates ground loops in multi-point systems. Standard construction recommendation for most industrial MIMS thermocouple applications. Slightly slower response than grounded due to the MgO thermal mass at the tip.
The thermocouple wires extend beyond the sheath tip and are welded together as a bare junction directly exposed to the measured medium. Provides the absolute fastest response time — typically T63 <100 ms in flowing gas. Not suitable for corrosive or pressurized processes. Used in gas temperature measurement, AMS 2750F TUS load sensors, and aerodynamic research probes where response speed is paramount.
Standard configurations available from stock or short lead time. Custom lengths, ODs, and terminations on request — no minimum order on custom dimensions. All ship with Certificate of Conformance and material certifications standard.
All types · MIMS construction · Inconel/Hastelloy sheaths · Class 1 tolerance · AS9100D documentation. Same-day response from engineering.