PT100 RTD Resistance Table
Complete PT100 and PT1000 resistance vs. temperature reference data per IEC 60751 (ITS-90), covering the full range from –200°C to +850°C. Includes Class AA, A, B, and C tolerance band data. Callendar-Van Dusen equation constants and interactive resistance calculator included.
Callendar-Van Dusen Equation — IEC 60751
The Callendar-Van Dusen equation is defined in IEC 60751:2022. Values are calculated on ITS-90 temperature scale. For T ≥ 0°C, C = 0 (term vanishes). The sensitivity (dR/dT) at 0°C is approximately 0.3851 Ω/°C for PT100 and 3.851 Ω/°C for PT1000.
PT100 Resistance Table — Sub-Zero Range (–200°C to 0°C)
Resistance values in ohms (Ω) at 1°C increments. Row = tens digit, column = units digit. Example: –45°C → row –40, col –5 = 82.29 Ω.
| Temp °C | 0 | –1 | –2 | –3 | –4 | –5 | –6 | –7 | –8 | –9 |
|---|---|---|---|---|---|---|---|---|---|---|
| –200 | 18.52 | — | — | — | — | — | — | — | — | — |
| –190 | 22.83 | 22.40 | 21.97 | 21.54 | 21.11 | 20.68 | 20.25 | 19.82 | 19.38 | 18.95 |
| –180 | 27.10 | 26.67 | 26.24 | 25.82 | 25.39 | 24.97 | 24.54 | 24.11 | 23.68 | 23.25 |
| –170 | 31.34 | 30.91 | 30.49 | 30.07 | 29.64 | 29.22 | 28.80 | 28.37 | 27.95 | 27.52 |
| –160 | 35.54 | 35.12 | 34.70 | 34.28 | 33.86 | 33.44 | 33.02 | 32.60 | 32.18 | 31.76 |
| –150 | 39.72 | 39.31 | 38.89 | 38.47 | 38.05 | 37.64 | 37.22 | 36.80 | 36.38 | 35.96 |
| –140 | 43.88 | 43.46 | 43.05 | 42.63 | 42.22 | 41.80 | 41.39 | 40.97 | 40.56 | 40.14 |
| –130 | 48.00 | 47.59 | 47.18 | 46.76 | 46.35 | 45.94 | 45.52 | 45.11 | 44.70 | 44.28 |
| –120 | 52.11 | 51.70 | 51.29 | 50.88 | 50.46 | 50.05 | 49.64 | 49.23 | 48.82 | 48.41 |
| –110 | 56.19 | 55.78 | 55.37 | 54.96 | 54.56 | 54.15 | 53.74 | 53.33 | 52.92 | 52.52 |
| –100 | 60.26 | 59.85 | 59.44 | 59.04 | 58.63 | 58.22 | 57.82 | 57.41 | 57.00 | 56.60 |
| –90 | 64.30 | 63.90 | 63.49 | 63.09 | 62.68 | 62.28 | 61.88 | 61.47 | 61.07 | 60.66 |
| –80 | 68.33 | 67.92 | 67.52 | 67.12 | 66.72 | 66.31 | 65.91 | 65.51 | 65.11 | 64.70 |
| –70 | 72.33 | 71.93 | 71.53 | 71.13 | 70.73 | 70.33 | 69.93 | 69.53 | 69.13 | 68.73 |
| –60 | 76.33 | 75.93 | 75.53 | 75.13 | 74.73 | 74.33 | 73.93 | 73.53 | 73.13 | 72.73 |
| –50 | 80.31 | 79.91 | 79.51 | 79.11 | 78.72 | 78.32 | 77.92 | 77.52 | 77.12 | 76.73 |
| –40 | 84.27 | 83.87 | 83.48 | 83.08 | 82.69 | 82.29 | 81.89 | 81.50 | 81.10 | 80.70 |
| –30 | 88.22 | 87.83 | 87.43 | 87.04 | 86.64 | 86.25 | 85.85 | 85.46 | 85.06 | 84.67 |
| –20 | 92.16 | 91.77 | 91.37 | 90.98 | 90.59 | 90.19 | 89.80 | 89.40 | 89.01 | 88.62 |
| –10 | 96.09 | 95.69 | 95.30 | 94.91 | 94.52 | 94.12 | 93.73 | 93.34 | 92.95 | 92.55 |
| 0 | 100.00 | 99.61 | 99.22 | 98.83 | 98.44 | 98.04 | 97.65 | 97.26 | 96.87 | 96.48 |
PT100 Resistance Table — Positive Range (0°C to +500°C)
Resistance values in ohms (Ω) at 1°C increments. Row = tens digit, column = units digit. Example: +123°C → row +120, col +3 = 147.28 Ω.
| Temp °C | +0 | +1 | +2 | +3 | +4 | +5 | +6 | +7 | +8 | +9 |
|---|---|---|---|---|---|---|---|---|---|---|
| 0 | 100.00 | 100.39 | 100.78 | 101.17 | 101.56 | 101.95 | 102.34 | 102.73 | 103.12 | 103.51 |
| 10 | 103.90 | 104.29 | 104.68 | 105.07 | 105.46 | 105.85 | 106.24 | 106.63 | 107.02 | 107.40 |
| 20 | 107.79 | 108.18 | 108.57 | 108.96 | 109.35 | 109.73 | 110.12 | 110.51 | 110.90 | 111.29 |
| 30 | 111.67 | 112.06 | 112.45 | 112.83 | 113.22 | 113.61 | 114.00 | 114.38 | 114.77 | 115.15 |
| 40 | 115.54 | 115.93 | 116.31 | 116.70 | 117.08 | 117.47 | 117.86 | 118.24 | 118.63 | 119.01 |
| 50 | 119.40 | 119.78 | 120.17 | 120.55 | 120.94 | 121.32 | 121.71 | 122.09 | 122.47 | 122.86 |
| 60 | 123.24 | 123.63 | 124.01 | 124.39 | 124.78 | 125.16 | 125.54 | 125.93 | 126.31 | 126.69 |
| 70 | 127.08 | 127.46 | 127.84 | 128.22 | 128.61 | 128.99 | 129.37 | 129.75 | 130.13 | 130.52 |
| 80 | 130.90 | 131.28 | 131.66 | 132.04 | 132.42 | 132.80 | 133.18 | 133.57 | 133.95 | 134.33 |
| 90 | 134.71 | 135.09 | 135.47 | 135.85 | 136.23 | 136.61 | 136.99 | 137.37 | 137.75 | 138.13 |
| 100 | 138.51 | 138.88 | 139.26 | 139.64 | 140.02 | 140.40 | 140.78 | 141.16 | 141.54 | 141.91 |
| 110 | 142.29 | 142.67 | 143.05 | 143.43 | 143.80 | 144.18 | 144.56 | 144.94 | 145.31 | 145.69 |
| 120 | 146.07 | 146.44 | 146.82 | 147.20 | 147.57 | 147.95 | 148.33 | 148.70 | 149.08 | 149.46 |
| 130 | 149.83 | 150.21 | 150.58 | 150.96 | 151.33 | 151.71 | 152.08 | 152.46 | 152.83 | 153.21 |
| 140 | 153.58 | 153.96 | 154.33 | 154.71 | 155.08 | 155.46 | 155.83 | 156.20 | 156.58 | 156.95 |
| 150 | 157.33 | 157.70 | 158.07 | 158.45 | 158.82 | 159.19 | 159.56 | 159.94 | 160.31 | 160.68 |
| 200 | 175.86 | 176.23 | 176.60 | 176.97 | 177.34 | 177.71 | 178.08 | 178.45 | 178.82 | 179.19 |
| 250 | 194.10 | 194.47 | 194.83 | 195.20 | 195.57 | 195.93 | 196.30 | 196.67 | 197.03 | 197.40 |
| 300 | 212.05 | 212.42 | 212.78 | 213.14 | 213.51 | 213.87 | 214.24 | 214.60 | 214.96 | 215.33 |
| 350 | 229.72 | 230.08 | 230.44 | 230.80 | 231.17 | 231.53 | 231.89 | 232.25 | 232.61 | 232.97 |
| 400 | 247.09 | 247.45 | 247.81 | 248.17 | 248.53 | 248.89 | 249.25 | 249.61 | 249.97 | 250.33 |
| 450 | 264.18 | 264.54 | 264.90 | 265.26 | 265.61 | 265.97 | 266.33 | 266.68 | 267.04 | 267.40 |
| 500 | 280.98 | 281.34 | 281.69 | 282.05 | 282.40 | 282.76 | 283.11 | 283.47 | 283.82 | 284.18 |
PT100 Resistance Table — High Range (+500°C to +850°C)
The PT100 operating range extends to +850°C per IEC 60751. Above +600°C, platinum element stability requires sheath protection from reducing atmospheres and contamination. Sensitivity (dR/dT) decreases slightly at high temperatures due to the quadratic B·T² term.
| Temp °C | +0 | +10 | +20 | +30 | +40 | +50 | +60 | +70 | +80 | +90 |
|---|---|---|---|---|---|---|---|---|---|---|
| 500 | 280.98 | 284.53 | 288.09 | 291.64 | 295.18 | 298.72 | 302.26 | 305.79 | 309.31 | 312.83 |
| 600 | 316.35 | 319.86 | 323.37 | 326.87 | 330.36 | 333.85 | 337.33 | 340.81 | 344.28 | 347.74 |
| 700 | 351.20 | 354.65 | 358.09 | 361.53 | 364.96 | 368.38 | 371.80 | 375.21 | 378.61 | 382.01 |
| 800 | 385.40 | 388.78 | 392.15 | 395.52 | 398.88 | 402.23 | 405.57 | 408.90 | 412.23 | 415.55 |
| 850 | 418.86 | — | — | — | — | — | — | — | — | — |
The IEC 60751 standard specifies that a 100 Ω platinum RTD at 0°C shall have a resistance of 418.68 Ω at +850°C — confirming the sensitivity relationship. For applications above 600°C, a Callendar-Van Dusen calculation with instrument linearization is recommended over table lookup due to the non-linearity increase.
PT1000 Resistance Table — Key Values
PT1000 resistance values are exactly 10× the PT100 values — R₀ = 1000 Ω at 0°C. PT1000 is preferred in low-power circuits and applications with long cable runs where 2-wire connection is required (the higher resistance minimizes the lead error fraction).
| Temperature | PT1000 (Ω) | PT100 (Ω) | Notes |
|---|---|---|---|
| –200°C | 185.2 | 18.52 | Cryogenic — LN₂ service |
| –100°C | 602.6 | 60.26 | Sub-zero industrial |
| –50°C | 803.1 | 80.31 | Freezer / cold storage |
| 0°C | 1000.0 | 100.00 | Ice point — calibration reference |
| +20°C | 1077.9 | 107.79 | Ambient reference |
| +37°C | 114.4 × 10 | 114.40 | Body temperature |
| +100°C | 1385.1 | 138.51 | Steam point — calibration |
| +200°C | 1758.6 | 175.86 | Process typical |
| +300°C | 2120.5 | 212.05 | High-temp industrial |
| +400°C | 2470.9 | 247.09 | High-temp furnace |
| +600°C | 3163.5 | 316.35 | Furnace / kiln |
| +850°C | 4188.6 | 418.86 | Upper limit IEC 60751 |
IEC 60751 Tolerance Classes — PT100 & PT1000
Tolerance is expressed as the greater of a fixed temperature value or a percentage of reading. The percentage term dominates at high temperatures. All Thermometrics RTDs are manufactured to IEC 60751 tolerance — Class AA through Class C available.
| Class | Tolerance Formula (±°C) | At 0°C | At 100°C | At 300°C | At 600°C | Applicability Range |
|---|---|---|---|---|---|---|
| AA (1/10 DIN) | ±(0.10 + 0.0017·|t|) | ±0.100 | ±0.270 | ±0.610 | ±1.120 | –50°C to +250°C |
| A | ±(0.15 + 0.002·|t|) | ±0.150 | ±0.350 | ±0.750 | ±1.350 | –100°C to +450°C |
| B | ±(0.30 + 0.005·|t|) | ±0.300 | ±0.800 | ±1.800 | ±3.300 | –196°C to +600°C |
| C (1/3 DIN) | ±(0.60 + 0.010·|t|) | ±0.600 | ±1.600 | ±3.600 | ±6.600 | –196°C to +600°C |
|t| = absolute value of temperature in °C. Class AA (also written 1/10 DIN or W0.1) is the tightest standard class. Custom tighter tolerances are available by special order on selected element types. Contact engineering for sub-Class AA requirements.