Precision pressure-containing thermowells per ASME B40.9 for RTD sensors and thermocouple assemblies. Drilled bar and fabricated construction in 316 SS, 316L, Inconel 600/625, and Hastelloy C-276. Straight, tapered, and stepped shank profiles. Threaded, flanged, and Van Stone connections.
A thermowell is a closed-end pressure-tight tube, permanently installed in a process pipe or vessel, into which a temperature sensor is inserted. The thermowell separates the sensor from the process fluid — protecting it against corrosion, erosion, flow velocity, and pressure — while allowing the sensor to be removed and replaced without taking the process offline or isolating pressure.
Thermowells are pressure-containing components and must be specified with adequate wall thickness for the rated pressure and temperature. They are also subject to vortex-induced vibration from process flow — requiring wake frequency analysis per ASME PTC 19.3 TW to ensure the natural frequency of the thermowell stem exceeds the vortex shedding frequency at maximum flow velocity.
Thermometrics manufactures thermowells per ASME B40.9 in all standard construction types, materials, and process connections. Matched thermowell-sensor assemblies (sensor + thermowell specified together) ensure correct bore diameter, immersion length, and lag characteristics for your application.
ASME B40.9 recognises two primary thermowell construction methods. Selection is governed by process conditions.
Manufactured from solid bar stock — the bore is produced by gun-drilling or deep-hole drilling through the full length. No weld seams exist in the pressure zone. Wall thickness is uniform and controlled throughout the immersion length. This construction provides the highest structural integrity and is mandatory per ASME B40.9 for high-pressure service.
Assembled by welding a precision-machined tip onto a tube shank. The tube provides the bore; the tip forms the closed end. Weld integrity is a critical quality characteristic — Thermometrics inspects all fabricated thermowell welds visually and by dye penetrant (PT) per ASME B40.9 requirements. Lower material cost and more economical for longer immersion lengths in lower-pressure service.
Shank profile determines the thermowell's wake frequency response, strength, and flow-induced stress characteristics. Selection is part of the ASME PTC 19.3 TW wake frequency calculation.
| Connection Type | Construction | Typical Pressure Rating | Application |
|---|---|---|---|
| Threaded — NPT | External male NPT thread (1/2" to 2" NPT). Screwed into a coupling or boss welded on pipe. | Up to Class 600 (1480 psi at 100°F for 316 SS) — pressure/temp rating per ASME B1.20.1 | Most common. General industrial piping. Easy retrofit installation. |
| Flanged — ASME B16.5 | Thermowell welded to flange matching pipe flange rating. Class 150 through Class 2500 available. | Class 150 to Class 2500 — full flange pressure class rating | High-pressure pipelines, vessels. Allows removal without breaking process connection. |
| Van Stone (Lap Joint) | Thermowell with loose lapped flange (field-rotatable). Thermowell tube slips through flange bore. | Same as flanged class selected | Where bolt-hole alignment must be field-adjusted. Cryogenic service (no socket weld thermal stress). |
| Weld-In / Socket Weld | Thermowell root machined with socket or butt weld prep. Welded directly into pipe nozzle. | Limited by pipe schedule — no connection fitting rating loss | Critical high-pressure service. No threaded connection leak path. |
| Sanitary Tri-Clamp | Sanitary clamp ferrule. 304 or 316L. Ra ≤ 0.8 µm bore finish standard. | Up to 150 psi typical | Pharmaceutical, food & beverage, biotech. 3-A compliant available. |
Material selection governs corrosion resistance, maximum operating temperature, and pressure rating. Thermometrics holds bar stock in all standard grades for prompt delivery.
Standard ASME B40.9 dimensions. Custom immersion lengths, bore diameters, and shank profiles available to order. All dimensions in millimetres (mm) unless noted.
| Model | Shank Type | Process Connection | Root OD (mm) | Tip OD (mm) | Bore Ø (mm) | Standard U (mm) | Construction |
|---|---|---|---|---|---|---|---|
| TW-S-NPT-100 | Straight | 1/2" NPT | 16.0 | 16.0 | 6.35 | 100, 150, 200, 250 | Drilled Bar |
| TW-S-NPT-150 | Straight | 3/4" NPT | 19.0 | 19.0 | 6.35 | 100, 150, 200, 250, 300 | Drilled Bar |
| TW-T-NPT-150 | Tapered | 3/4" NPT | 25.4 | 12.7 | 6.35 | 150, 200, 250, 300, 400 | Drilled Bar |
| TW-T-NPT-200 | Tapered | 1" NPT | 31.8 | 15.9 | 6.35 | 200, 250, 300, 400, 500 | Drilled Bar |
| TW-ST-NPT-150 | Stepped | 3/4" NPT | 22.2 | 15.9 | 6.35 | 150, 200, 250, 300 | Drilled Bar |
| TW-T-FL150-300 | Tapered | ASME B16.5 CL150 RF | 25.4 | 12.7 | 6.35 | 200, 300, 400, 500 | Drilled Bar |
| TW-T-FL300-300 | Tapered | ASME B16.5 CL300 RF | 31.8 | 15.9 | 6.35 | 200, 300, 400, 500 | Drilled Bar |
| TW-S-TC-100 | Straight | Tri-Clamp 1.5" | 16.0 | 16.0 | 6.35 | 100, 150, 200 | Fabricated |
| TW-S-TC-150 | Straight | Tri-Clamp 2" | 16.0 | 16.0 | 6.35 | 150, 200, 250 | Fabricated |
| Custom immersion lengths, bore diameters (6.35 mm standard; 4.76 mm and 8.0 mm available), and materials to order. U-lengths above are standard stock; non-standard cut to length with 3–5 day lead time. | |||||||
Flow past a thermowell shank produces alternating vortices — the von Kármán vortex street — at a frequency determined by the Strouhal number, process flow velocity, and shank tip diameter. This is vortex shedding.
If the vortex shedding frequency approaches the thermowell's natural frequency (determined by stem stiffness, length, and tip mass), resonance occurs. Resonant vortex-induced vibration produces cyclic bending stress at the thermowell root that can cause fatigue failure — the single most common thermowell failure mode in process plants.
ASME PTC 19.3 TW (2016) provides the governing calculation procedure. The Strouhal frequency ratio (rf = fS/fn) must remain below 0.8 under maximum flow velocity conditions with a defined safety factor. When the ratio cannot be reduced by shank profile selection alone, shortened immersion length, increased shank OD, or alternative mounting (isolation chamber or thermowell cage) may be required.
Thermometrics can perform or independently review wake frequency calculations on request. Provide process fluid density, dynamic viscosity, maximum flow velocity, pipe internal diameter, and proposed thermowell dimensions.