IEC 60584 Type R — Thermocouple Data
Positive LegPlatinum-13% Rhodium (Pt-13%Rh)
Negative LegPlatinum (Pt)
Full Range–50°C to +1768°C
Practical Range+300°C to +1600°C practical
Seebeck Coefficient~10.5 µV/°C at 1000°C; ~7.5 µV/°C at 300°C
Class 1 Tolerance±1.0°C or ±0.25%|t|
Class 2 Tolerance±1.5°C or ±0.25%|t|
Class 1 IEC Range+600°C to +1600°C
Class 2 IEC Range+600°C to +1600°C
Color Code (IEC)Orange (+) / White (–)
Color Code (ANSI)Black (+) / Red (–)
Standard SheathHigh-purity alumina ceramic protection tube
ConstructionMIMS (standard) or conventional; ungrounded, grounded, or exposed junction
DocumentationAS9100D COC + material certs standard; NIST-traceable cal cert on request
Application Notes
Type R is a noble metal thermocouple offering the highest accuracy and long-term stability of any IEC-standardized type in oxidizing atmospheres above 600°C. The IEC Class 1 tolerance of ±1.0°C at 1000°C is significantly tighter than any base-metal type. NIST-traceable calibration certificates with expanded uncertainty statements (k=2) are standard on all Type R shipments from Thermometrics. For AMS 2750F Nadcap compliance, Type R meets the requirements for SAT reference sensors.