IEC 60584 Type N — Thermocouple Data
Positive LegNicrosil (Ni-14%Cr-1.4%Si)
Negative LegNisil (Ni-4.4%Si-0.1%Mg)
Full Range–270°C to +1300°C
Practical Range–200°C to +1260°C
Seebeck Coefficient~36.7 µV/°C at 25°C; ~38.9 µV/°C at 600°C
Class 1 Tolerance±1.5°C or ±0.4%|t|
Class 2 Tolerance±2.5°C or ±0.75%|t|
Class 1 IEC Range–40°C to +1000°C
Class 2 IEC Range–40°C to +1200°C
Color Code (IEC)Pink (+) / White (–)
Color Code (ANSI)Orange (+) / Red (–)
Standard SheathInconel 600 standard; Inconel 625 for extended service
ConstructionMIMS (standard) or conventional; ungrounded, grounded, or exposed junction
DocumentationAS9100D COC + material certs standard; NIST-traceable cal cert on request
Application Notes
Type N was specifically engineered to eliminate the drift mechanisms present in Type K. Silicon additions to both the Nicrosil (+) and Nisil (–) legs stabilize the alloy microstructure and suppress the ordering transformation responsible for K-effect drift between 200–650°C. The result is 3–10× less drift than Type K over equivalent service life. For AMS 2750F Nadcap heat treating facilities seeking to extend TUS and SAT calibration intervals, Type N is the preferred choice for control and load sensors.