IEC 60584 Type K — Thermocouple Data
Positive LegChromel (Ni-10%Cr)
Negative LegAlumel (Ni-5%Al-2%Si)
Full Range–270°C to +1372°C
Practical Range–200°C to +1260°C
Seebeck Coefficient~40.6 µV/°C at 25°C
Class 1 Tolerance±1.5°C or ±0.4%|t|
Class 2 Tolerance±2.5°C or ±0.75%|t|
Class 1 IEC Range–40°C to +1000°C
Class 2 IEC Range–40°C to +1200°C
Color Code (IEC)Green (+) / White (–)
Color Code (ANSI)Yellow (+) / Red (–)
Standard SheathInconel 600 recommended above 500°C
ConstructionMIMS (standard) or conventional; ungrounded, grounded, or exposed junction
DocumentationAS9100D COC + material certs standard; NIST-traceable cal cert on request
Application Notes
Most widely used thermocouple type globally. Good linearity and wide instrument support. Susceptible to short-range ordering (K-effect) between 200–650°C causing reversible drift of up to ±5°C in long-term furnace service — specify Type N for applications requiring superior stability in this range. Positive leg is magnetic. Not recommended for sulfurous, reducing, or vacuum atmospheres without protection tube.