IEC 60584 Type T — Thermocouple Data
Positive LegCopper (Cu)
Negative LegConstantan (Cu-45%Ni)
Full Range–270°C to +400°C
Practical Range–200°C to +350°C
Seebeck Coefficient~40.9 µV/°C at 25°C; ~17 µV/°C at –200°C
Class 1 Tolerance±0.5°C or ±0.4%|t| (absolute ±0.5°C below 125°C)
Class 2 Tolerance±1.0°C or ±0.75%|t|
Class 1 IEC Range–40°C to +125°C
Class 2 IEC Range–40°C to +350°C
Color Code (IEC)Brown (+) / White (–)
Color Code (ANSI)Blue (+) / Red (–)
Standard Sheath316L SS standard; copper/Teflon for pharma and food
ConstructionMIMS (standard) or conventional; ungrounded, grounded, or exposed junction
DocumentationAS9100D COC + material certs standard; NIST-traceable cal cert on request
Application Notes
Type T is the preferred thermocouple for sub-zero and cryogenic applications. Both copper and constantan legs resist moisture corrosion, making Type T suitable for wet or humid environments. The IEC Class 1 absolute tolerance of ±0.5°C between –40°C and +125°C is the tightest fixed tolerance of any IEC base-metal thermocouple type. Copper positive leg limits maximum temperature to +400°C. NIST-traceable calibration at sub-zero setpoints is standard at Thermometrics for pharmaceutical and lyophilizer validation applications.